3-20 M1207026
Remove the test lung from the patient connection.
Other alarms such low minute volume can occur.
Make sure that the low airway pressure and apnea alarms
occur. The apnea alarm occurs after 15 seconds.
12. Test the sustained airway pressure alarm:
Set the controls:
– APL valve: Closed (70 cm H
2
O)
Mechanical ventilation stops when the Bag/Vent switch is set
to Bag.
Occlude the bag port connector with a test plug.
Close the patient connection using the test plug located on
the side of patient Y-piece connector and push the O
2
flush
button.
Make sure that the sustained pressure alarm occurs after
approximately 15 seconds at the sustained pressure limit (10
cm H
2
O).
Note If the system does not include O
2
monitor, skip step 13 and 14.
13. Test the O
2
monitor and alarms
Remove the O
2
sensor from the breathing system.
Make sure the sensor measures approximately 21% O
2
in
room air.
Set the low O
2
alarm to 50%. Make sure a low O
2
alarm
occurs.
Set the low O
2
alarm back to 21% and make sure that alarm
cancels.
Put the O
2
sensor back in circuit.
Remove the test lung from the patient connection.
Set the High O
2
alarm to 50%.
Push the O
2
flush button to fill the breathing system.
Set the O
2
flow control to 5 L/min.
Make sure the high O
2
alarm comes On.
Set the high O
2
alarm back to 100% and make sure that
alarm cancels.
After 2 minutes in pure O
2
, the O
2
display reads
approximately 100%.
Turn the O
2
flow control fully clockwise (zero flow).
14. Set the system switch to Standby.
M1207027 004 MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX