GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
7-16 Section 7-5 - Remote Access to Service Platform
The screen below confirms the Disruptive Mode status change.
7.) After “Disruptive Mode” is enabled, “Virtual Console Observation” can be activated by selecting
Utilities -> Common Utilities -> Virtual Console Observation, and then pressing the START
button.
NOTE: An attempt to activate the Virtual Console Observation Tool without first enabling “Disruptive Mode” is
impossible.
If “Virtual Console Observation” has been activated successfully, following display is shown to the FE.
After “Virtual Console Observation” is established, the Voluson® P8 / Voluson® P6 which should be
observed, can be accessed.
By means of the “Virtual Network Computing (VNC)” feature (see: Section 7-5-5 on page 7-21) the
Service technician can operate on the customer’s PC using the mouse and the keyboard.
NOTE: To reduce the transmission time the colors are reduced.
Figure 7-17 Disruptive Mode enabled
Figure 7-18 Virtual Console Observation is running