GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013)  PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-31
Section 7-8      How to repair Databases (*.mdb) (cont’d)
8.) After terminating all processes mentioned above, press CTRL
 + ESC to open windows start menu.
9.) Using the [Arrow] keys on the keyboard, select “Programs” -> “GE FIELD ENGINEER” and then 
double-click the “Compact and Repair Database” application.
It reports on the actions it performs. This can be valuable information for the technicians. Check that all 
databases (.mdb) are repaired (see:Figure 7-37).
10.)In case of any failures, please make a screen shot (press ALT
 + P key, to save it to D:\export), print 
out or write down the error message and then close the program by pressing ENTER
.
11.)Close all open windows (e.g., Windows Explorer).
12.)Restart the system (turn off and on the system).
Figure 7-37  double-click “CheckDB.exe”
Figure 7-38  Report repairing Databases