EasyManuals Logo

HBK GEN7iB GEN Series User Manual

HBK GEN7iB GEN Series
519 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #203 background imageLoading...
Page #203 background image
GEN7iB
203
INPUT CARDS
INPUT CARDS
12
12.2.9 GN310B/GN311B Isolation Current channel
Thespecied±20VDCvoltagerangeoftheCurrentchannelinputsinvoltagemodeis
suchthatitfallsbelowthelowvoltagelimitasspeciedinIEC61010-1.
The isolation of the Current channel inputs is in line with the limit for safe voltage and
currents as mentioned above.
Current Channels Isolation IEC 61010-2-030:2017
Ͳ
н
$'&
,VRODWHGFKDQQHO
Ͳ
н
$'&
,VRODWHGFKDQQHO
9'&
9506
9'&
9506
9'&
9506
&KDVVLV
Fig. 12.26 Current channels isolation ratings
Current Channels Isolation
Input pin to chassis ± 60 V DC, 30 V RMS
Channel to channel 120 V DC, 50 V RMS
The input channels of the GN310B and GN311B are isolated. This means that the
inputchannelandamplierarefullyisolatedfrom(earth)ground.Inthiscontext,fully
isolated means a very high resistance and a very small capacitive coupling to ground.
This is for safety and to avoid ground loops.