FREQUENTLY ASKED QUESTIONS
FREQUENTLY ASKED QUESTIONSE
E Frequently Asked Questions
E.1 Overview
GEN series Trigger Arm
Question: Answer:
Can Trigger Arm be used in a multi main-
frame setup using Master/Sync connec-
tion?
Yes, Trigger Arm can be used in a multi
mainframe setup using Master/ Sync
connection when the arm signal is
provided to all mainframes like the TTL
signal External Event In or the CAN bus.
TriggerArmusesthequalierlinesto
allow triggering or not.
Master/Syncsetupsharesthequalier
and trigger lines over the multi Main-
frames. In case one mainframe isn’t being
armed the whole multi Mainframe setup
cannottriggerduetothequalierstate.
GEN series External Trigger Out delay
Question: Answer:
Is there a difference in External Trigger
Out delay from a calculated (RT-FDB)
channel and input channel trigger detec-
tor?
Yes, External Trigger Out delay from a cal-
culated channel will always have a delay
of 100 ms. The External Trigger Out delay
from an input channel is user selectable
between10μsand516μs.
Touch screen disinfection
Question: Answer:
How do I clean and disinfect the touch
screen properly during COVID-19?
HBM only uses ELO touch screens. For
information on the correct cleaning of the
touch screen, please refer to the manu-
facturer's website:
www.elotouch.com/support/technical-
support/cleaning
Note: In case this link fails, search the
ELOTOUCH website for COVID-19 support.