UNDERSTANDING INPUTS AND USAGE OF PROBES
UNDERSTANDING INPUTS AND USAGE OF PROBES
15
Differentampliers–ProsandCons
Single-ended(toground)ampliers
Cost effective and small
High bandwidth
Easy to use with probes
Potential ground problems
No safety problems
No CMRR and no CMV
Single-endedisolatedamplier–unbalanceddifferential
Can perform differential measurements
Expensive and large
Diculttousewithprobes
Limited CMRR, best CMV
Avoids ground loops
High level of safety
Differentialamplier(withcommonground)
Widely used in DAQ
Good CMRR, limited CMV
No (safety) isolation and potential ground loops will remain present
Differentialamplierwithisolatedcommon
Safe
Expensive and large
Good CMRR, best CMV
Morediculttousewithprobes