9.4 Command
______________________________________________________________________________________________
Syntax
:SYS:OPT?
Response
syntax
<data1><data2><data3><data4><data5><data6><data7><data8><data9><data10><data11>
<data>
<data1> PASS hold function
0: Not held, 1: Held
<data2> FAIL hold function
0: Not held, 1: Held
<data3> Hold function
0: Not held, 1: Held
<data4> Momentary out function
0: Not set, 1: Set
<data5> Double Action
0: Not set, 1: Set
<data6> FAIL mode
0: Not set, 1: Set
<data7> RS command [START] setting
0: Not set, 1: Set
<data8> Inter-lock function
0: Not set, 1: Set
<data9> Voltage comparator position
0: Start test, 1: End test
<data10> Insulation resistance test measurement range
0: Fixed range, 1: Auto range
<data11> Insulation resistance test termination mode
0: Test for set time, 1: Terminate test at PASS screening
2: Terminate test at FAIL screening
Function
Queries the optional function settings.
Example
Queries the optional function settings.
Transmission :SYS:OPT?
Queries the optional function settings.
Response 01000010001
You cannot make settings other than the following: FAIL Hold function, RS
[START] command settings, end the insulation resistance test at PASS
screening.
:SYS:OPT?
Queries the optional function settings.