9.4 Command
______________________________________________________________________________________________
Syntax
:SYS:OPT <data1><data2><data3><data4><data5><data6><data7><data8><data9>
<data10><data11>
<data>
<data1> PASS hold function
0: Not held, 1: Held
<data2> FAIL hold function
0: Not held, 1: Held
<data3> Hold function
0: Not held, 1: Held
<data4> Momentary out function
0: Not set, 1: Set
<data5> Double Action
0: Not set, 1: Set
<data6> FAIL mode
0: Not set, 1: Set
<data7> RS command [START] setting
0: Not set, 1: Set
<data8> Inter-lock function
0: Not set, 1: Set
<data9> Voltage comparator position
0: Start test, 1: End test
<data10> Insulation resistance test measurement range
0: Fixed range, 1: Auto range
<data11> Insulation resistance test termination mode
0: Test for set time, 1: Terminate test at PASS screening
2: Terminate test at FAIL screening
Response
syntax
OK
Function
Sets the optional functions.
Example
Sets the optional functions.
Transmission :SYS:OPT 01000010001
Do not make settings other than the following: FAIL Hold function,
RS [START] command settings, end the insulation resistance test at PASS
screening.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when a value that is not set is specified.
:SYS:OPT
Sets the optional functions.