EasyManua.ls Logo

Hioki 3159 - Chapter 11 Maintenance and Inspection

Hioki 3159
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
149
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:MEM:INS:SAVE <data>
<data>
1 to 10 (NR1)
Response
syntax
OK
Function
Saves the current insulation resistance test settings in the memory file specified in <data>.
Example
Saves insulation resistance test data to memory files.
Transmission :MEM:INS:SAVE 2
Saves the insulation resistance test settings in Memory File 2.
Response OK
Save completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
MEM:INS:CLE <data>
<data>
1 to 10 (NR1)
Response
syntax
OK
Function
Clears and resets the data saved for the insulation resistance test specified in <data>.
Example
Clears the insulation resistance test data saved to memory files.
Transmission :MEM:INS:CLE 3
Clears and resets the content of insulation resistance test Memory File 3.
Response OK
Clear completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
:MEM:INS:SAVE
Saves data for insulation resistance tests to memory files.
:MEM:INS:CLE
Clears the insulation resistance test data saved to memory files.