9.4 Command
______________________________________________________________________________________________
Syntax
:MEM:INS:FILE? <data>
<data>
1 to 10 (NR1)
Response
syntax
Test voltage value 0.00 to 5.00 (NR2)
Upper limit value 0.2 to 2000 (NR1 or NR2)
Lower limit value 0.2 to 2000 (NR1 or NR2)
Test time setting value 0.5 to 999 (NR1 or NR2)
When each setting is set to off, "0" will be returned.
Function
Queries the content of saved data for the insulation resistance file specified in <data>.
Example
Queries the data saved for insulation resistance tests.
Transmission :MEM:INS:FILE? 1
Queries the insulation resistance test Memory File 1 settings.
Response 1000,0,100.0,5.0
Test voltage: 1000 V; Upper-limit value: OFF;
Lower-limit value: 100.0 M; Test time: 5.0 s
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
:MEM:INS:LOAD <data>
<data>
1 to 10 (NR1)
Response
syntax
OK
Function
Loads the data saved for the insulation resistance file specified in <data>.
Example
Loads data saved for insulation resistance tests.
Transmission :MEM:INS:LOAD 1
Loads the insulation resistance test Memory File 1.
Response OK
Loading completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
:MEM:INS:FILE?
Queries the content of data saved for insulation resistance tests.
:MEM:INS:LOAD
Loads data saved for insulation resistance tests.