147
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:MEM:WITH:SAVE <data>
<data>
1 to 10 (NR1)
Response
syntax
OK
Function
Saves the current withstand voltage test settings in the memory file specified in <data>.
Example
Saves data for withstand voltage tests to memory files.
Transmission :MEM:WITH:SAVE 2
Saves the withstand voltage test settings in Memory File 2.
Response OK
Save completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
:MEM:WITH:CLE <data>
<data>
1 to 10 (NR1)
Response
syntax
OK
Function
Clears and resets the data saved for the withstand voltage test specified in <data>.
Example
Clears the withstand voltage test data saved to memory files.
Transmission :MEM:WITH:CLE 3
Clears and resets the content of withstand voltage test Memory File 3.
Response OK
Clear completion
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
:MEM:WITH:SAVE
Saves data for withstand voltage tests to memory files.
:MEM:WITH:CLE
Clears the withstand voltage test data saved to memory files.