3.5 PASS or FAIL Determination
______________________________________________________________________________________________



NOTE
If the optional "Voltage Comparator Position" function is set to "
1
:Endoftest
time", since the unit shifts to the TEST state whenever voltage is output,
TEST
does not flash before the output-voltage value reaches the comparative-voltage
value range.
If the optional "Voltage Comparator Position" function is set to "
1
:Endoftest
time", this function checks whether the output voltage is within the standard
volta
g
eran
g
e onl
y
when the test is terminated.
When the output-voltage deviates from the comparative-voltage range:
1.
Press the START key to start a test.
2.
TEST
will flicker until the output voltage switches to the
comparativevoltage range.
3.
If the output voltage fails to switch to this range in 5 seconds, the unit
switches to the FAIL state and stops outputting a voltage.
4. In the FAIL state,
UPPER
,
LOWER
,and
FAIL
light up to indicate that an
improper voltage has been output.