42
_____________________________________________________________________________________________
3.5 PASS or FAIL Determination
______________________________________________________________________________________________
FAIL Screening Using the Comparative-Voltage Value
When the output-voltage value fails to reach the comparative-voltage range:
1. Press the
START key to start a test.
2.
TEST
will flicker until the output voltage switches to the
comparativevoltage range.
3. If the output voltage fails to switch to this range in 5 seconds, the unit
switches to the FAIL state and stops outputting a voltage.
4. In the FAIL state,
UPPER
,
LOWER
,and
FAIL
light up to indicate that an
improper voltage has been output.