4.3 "SETTING " State
______________________________________________________________________________________________
In this example, the value is 1000 V.
Using the / keys, set the comparative voltage value to 500 V.
(3) Setting an lower-limit value
Press the key to move the flashing cursor to the lower-limit value.
In this example, switch from 10.0 MΩ to 100 MΩ using the / keys.
To change the lower limit value, press / keys.
Press the keys, SHIFT + / key. The value increases or decreases 10 times
more rapidly than when the SHIFT key is not held down.
(4) Setting a upper-limit value
Using the key, move the flashing cursor to the upper limit value.
The upper-limit value is set at 2000 MΩ. Turn it OFF, as it is not needed. To
change to OFF, press the ON/OFF key.