57
_____________________________________________________________________________________________
4.3 "SETTING " State
______________________________________________________________________________________________
Test voltage value: 500 V
Upper limit value: 100 M
Lower limit value: OFF
Test time: 5.0 s
(5) Setting the test time
Using the key, move the flashing cursor to the test time.
In this example, change the test time from
20.0 s
to
5.0 s
.
(6) Changing to the READY state
To return to the READY state, press the STOP key, which will finalize the
test settings. In the READY state, the displayed test-voltage value is replaced
with a measured-voltage value, and
READY
lights up. The test-voltage value is
shown below.
The new parameters following setting are shown below: