92
────────────────────────────────────────────────────
8.3.3 Specific Commands for Output Settings
────────────────────────────────────────────────────
Syntax
:OSET:FREQuency? <ch>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
Response
syntax
(Header ON)
:OSET:FREQ <ch>,<frequency>
(Header OFF)
<ch>,<frequency>
Example
Transmission
Response
:OSET:FREQ? 1
(Header ON)
:OSET:FREQUENCY 1,100000.00
(Header OFF)
1,100000.00
Syntax
:OSET:PERiod <ch>,<period>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
<period>
Period of SIN and rectangular
waveforms
= 100 to 0.0000001 (seconds)
Clock period of an arbitrary
waveform
= 100 to 0.0000001 (seconds)
Period of waveform other than
custom
= 100 to 0.000005 (seconds)
Example
Transmission
:OSET:PER 1,1e-6
Sets the frequency for channel 1 to 1
MHz.
Function
Returns the frequency for the specified
channel as NR2 numerical data.
Note
With this query, if any error occurs, no
response message is produced.
Errors
(Execution error)
When DC or noize is selected as the
waveform type
When sweep is on
Function
Sets the frequency of a basic
waveform, or the clock frequency of a
arbitrary waveform.
Errors
(Execution error)
When the specified value cannot be
set
When DC or noize is selected as the
waveform type
When sweep is on
:OSET:FREQuency?
Queries the frequency.
:OSET:PERiod
Sets the period.

Table of Contents