93
────────────────────────────────────────────────────
8.3.3 Specific Commands for Output Settings
────────────────────────────────────────────────────
Syntax
:OSET:PERiod? <ch>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
Response
syntax
(Header ON)
:OSET:PER <ch>,<period>
(Header OFF)
<ch>,<period>
Example
Transmission
Response
:OSET:PER? 1
(Header ON)
:OSET:PERIOD 1,0.0000001
(Header OFF)
1,0.0000001
Syntax
:OSET:RANGe <ch>,<range>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
<range>
= R10V (10 V range)
= R1V (1 V range)
= R0_1V (0.1 V range)
Example
Transmission
:OSET:RANG 1,R1V
Sets the range for channel 1 to 1 V.
Syntax
:OSET:RANGe? <ch>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
Response
syntax
(Header ON)
:OSET:RANGE <ch>,<range>
(Header OFF)
<ch>,<range>
Example
Transmission
Response
:OSET:RANG? 1
(Header ON)
:OSET:RANGE 1,R1V
(Header OFF)
1,R1V
Function
Returns the period for the specified
channel as NR2 numerical data.
Note
With this query, if any error occurs, no
response message is produced.
Error
(Execution error)
When DC or noize is selected as the
waveform type
Function
Sets the range.
Error
(Command error)
When characters are specified other
than those at the left.
(Execution error)
When RUN is active.
Function
Returns the range for the specified
channel as a character data.
Note
With this query, if any error occurs, no
response message is produced.
:OSET:PERiod?
Queries the period.
:OSET:RANGe
Sets the range.
:OSET:RANGe?
Queries the range.