94
────────────────────────────────────────────────────
8.3.3 Specific Commands for Output Settings
────────────────────────────────────────────────────
Syntax
:OSET:DUTY <ch>,<duty>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
<duty> = 1.0 to 99.0 %
Example
Transmission
:OSET:DUTY 1,50
Sets the duty for channel 1 to 50%.
Syntax
:OSET:DUTY? <ch>
<ch> Channel number
= 1 to 4 (1 to 2 for the 7075-01)
Response
syntax
(Header ON)
:OSET:DUTY <ch>,<duty>
(Header OFF)
<ch>,<duty>
Example
Transmission
Response
:OSET:DUTY? 1
(Header ON)
:OSET:DUTY 1,50.0
(Header OFF)
1,50.0
Function
Sets the duty cycle of a pulse
waveform.
Errors
(Execution error)
When the specified value cannot be
set
When the specified waveform is not a
pulse waveform
When sweep is on
Function
Returns the duty for the specified
channel as NR2 numerical data.
Note
With this query, if any error occurs, no
response message is produced.
Errors
(Execution error)
When the specified waveform is not a
pulse waveform
When sweep is on
:OSET:DUTY
Sets the duty.
:OSET:DUTY?
Queries the duty.

Table of Contents