Table 16-12 Detailed Information about "SPE: Text for test selecting information inquiry (from analyzer to host)"
TS: Test selecting information
2 Characters Sample Information
Operation Instrument Sample 5 Characters 1 Character 3 Characters 13 Characters 15 Characters
Mode Mode Name Function
Character
Sample No. Unassigned Position No. ID No. Unassigned Remarks
Simple
analysis
mode
ID mode Routine
sample
Press the registration key on the
screen by the same times as the
sample count for inquiry.
Sample
unregistered
A 1 to 400 Space 1 to 35 ID read with
barcode reader
Space Even if a barcode read error occurs due
to an unregistered sample, inquiry is
made with position No. assigned.
Sample registered Same as
above
Inquiry is not made when [NO] is
specified for full-time inquiry and TS
request lasts 1 channel or more.
Stat
sample
When [RUN] (highlighted) is specified for stat sample
test selection, inquiry is made while reading barcode
from the position specified for stat sample position
setting parameter.
D 1 to 50 Space 2 to 35 Same as
above
Space Same as above
Sample No.
mode
Routine
sample
Press the registration key on the
screen by the same times as the
sample count for inquiry.
Sample
unregistered
N 1 to 400 Space 1 to 35 ID registered
on screen
[handled as
Space Same as above
Sample registered comment]
Stat
sample
Inquiry is not made.
Routine
analysis
mode
ID mode Routine
sample
Inquiry is made by reading
barcodes sequentially from
position No. 1.
Sample
unregistered
A Space Space 1 to 35 ID read with
barcode reader
Space Inquiry is not made when an
unregistered sample causes barcode
read error.
Sample registered Same as
above
Inquiry is not made when [NO] is
specified for full-time inquiry and TS
request lasts 1 channel or more.
Stat
When [RUN] (highlighted) is specified for stat sample
D
Space
Space
2 to35
ID registered
Space
Same as above
sample test selection, inquiry is made while reading barcode
from the position specified for stat sample position
setting parameter.
(position Nos.
read by
barcode
reader)
on screen
Sample No.
mode
Routine
sample
Inquiry is made for the specified
number of samples sequentially
starting from the sample
specified in the analysis start
Sample
unregistered
N 1 to 400 Space 1 to 35 ID registered
on screen
[handled as
comment]
Space Same as above
No. on start condition screen. Sample registered 1 to 400
Stat
sample
Inquiry is not made.
Supplementary Explanation
1. If no response is available from the host within 2 cycles (36 sec) after inquiry to the host, analysis is carried out according to the TS of main frame.
If the main frame has no TS request, the relevant sample will not be analyzed.
Inquiry Condition
[RUN] (highlighted) is specified for host
communication on start condition screen, [NO] is
specified for the transfer of analytical data alone on
the [communication parameter screen and [NO] is
specified for original Abs.
When barcode reading is
successful, inquiry is
made via the read ID.
When unsuccessful,
inquiry is made via the ID
specified on screen

Table of Contents