Waveform DisplayWaveform Display
Displayed Waveforms Eight waveform windows maximum, with scrolling across 96
waveforms.
Display Formats Waveforms can be displayed in an overlapping and/or
non-overlapping format. Display capability of A− B and A+B is
also pr ovided.
Display Resolution 500 points horizontal, 240 points vertical
Display ModesDisplay Modes
Normal (Single) New acquisition replaces old acquisition on screen.
Accumulate New acquisition is displayed in addition to previous
acquisitions until screen is erased.
Average New acquisitions are averaged wit h old acquisit ions, wit h
the updated waveform displayed until erased. Maximum
number of averages is 256.
Overlay Up to 8 acquired waveforms from separate channels can be
overlayed in the same display area.
Connect-the-dots Provides a display of the sample points connected by
straight lines.
Waveform Reconstruction When there is insufficient data to fill every horizontal
location, a post acquisition reconstruction filter fills in
missing data points for time base < 50 ns/div.
Waveform Math Display capability of A− B and A+B functions is provided.
General Information
Supplemental Characteristics (oscilloscope)
1–11