Keysight B1505A Configuration and Connection Guide 2-15
N1259A Connection Guide
Output Connection
Figure 2-6 Module Selector Simplified Internal Connections
NOTE Controlling module selector input-to-output path
The configuration and the default I/O path of the module selector are set by using the
Module Selector tab screen of the Configuration window of the EasyEXPERT software.
The connection path from the input port to the output port is controlled by each test setup.
And the module selector switching status is indicated by the status indicator mounted on
the front panel. See Figure 2-6 and Table 2-6 for the indication and the connection path.
Table 2-6 Module Selector Status Indicator and Input-to-Output Connection Path
Low High
SF
S: Sense
F: Force
P.A
P.A
GNDU HCSMU HPSMU HVSMU
SF
SF
SF
F
P.A:
Protection adapter
SF
GNDU force and sense
are connected to Low
sense line.
HVSMU force is connected
to High sense line.
100 kW
Module used for the test
Status
indicator
Input-to-output connection
1
of output terminals
1. In the normal state, the module selector makes the path to the output port from the input port specified
by the Default field. When a measurement is performed, the module selector performs automatic
switching in every test. The selector makes the path to the output port from the module used for the test.
Low Sense Low Force High Sense High Force
2
2. This condition is when all Input fields HVSMU, HCSMU, and HPSMU are blank.
Open HCSMU3
Low Sense
+
GNDU2 Force
+
GNDU2 Sense
HCSMU3
Low Force
Open Open
HPSMU or MPSMU HPSMU HPSMU3 Sense HPSMU3 Force
HCSMU or DHCSMU HCSMU HCSMU3
High Sense
HCSMU3
High Force
HVSMU or HVMCU HVSMU HVSMU2 Force
Open
HVSMU/HVMCU and
built-in series resistor
HVSMU2 Force
+ Series resistor