2-16 Keysight B1505A Configuration and Connection Guide
N1259A Connection Guide
Output Connection
To Use Series Resistor
The series resistor is used to reduce damage of DUT or prevent SMU from oscillation. The
resistor is internally connected between the terminals 1 and 2.
NOTE To install your desired resistor in the universal R-box, see "To Use Universal R-Box".
Required parts:
N1254A-508 or N1254A-509 connection wire, 2 ea., or 3 ea. for Kelvin connection
Connection:
See Figure 2-7.
1. Connect a wire between a SMU Force output terminal and the 1 terminal.
2. Connect a wire between the 2 terminal and a Force terminal on the socket module.
This is the non-Kelvin connection, solid lines shown in Figure 2-7. For the Kelvin
connections, add the following connection, dotted line shown in Figure 2-7.
1. Connect a wire between the SMU Sense output terminal and the 1 terminal.
The Kelvin connection is effective up to the series resistor input. So the voltage drop
caused by the series resistor must be subtracted from the measurement result as
following.
Vdut = Vout - Rs Imeas
where,
Vdut: Voltage after correction
Vout: Source output voltage
Rs: Resistance of the series resistor
Imeas: Measurement current
Figure 2-7 Series Resistor Connection Example
High Voltage Bias–Tee
Guard
Module Selector Output
Force
Sense
Guard
Sense
Force
1 MΩ
21
100 kΩ
HighLow
21
±3 kV Max±40 V Max
N1259A
Opt 022
±3 kV Max
Max: 6.4 W
±3 kV Max
Max: 9 W
N1259A
Opt 021
N1259A
Opt 010
Force
Sense
Force
Sense
Force
Sense
1
2
3
123
±3 kV Max
Inline Package Socket (3 Pin)
Kelvin Connection
MF CMU
DC Bias Input
±3 kV Max
High
Low
Force
±3 kV Max
±25 V Max
Guard
Sense
Force

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals