2 General Power Meter Functions
108 Keysight N1913/1914A User’s Guide
Keysight 8485D Prior to the power meter being zeroed and calibrated, a Keysight 11708A 30 dB reference
attenuator and an APC 3.5 (f) to 50 Ω (m) N-Type adapter (08485-60005) should be
connected between the power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Keysight 8487A This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter (08487-60001) to
connect to the power meter. Remove this attenuator before making measurements.
Keysight 8487D Prior to the power meter being zeroed and calibrated, a Keysight 11708A 30 dB reference
attenuator and an APC 2.4 (f) to 50 Ω (m) N-Type adapter (08487-60001) should be
connected between the power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Sensor Model Connection Requirements