General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 71
Selectable Frequency Dependent Offset Unit (dB or %)
The offset in dB range is from –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
Once the offset unit is selected, it will be applied to all the offsets in the selected
table and also to the remaining nine tables.
To change the offset unit to dB, press , , and .
To change the offset unit to %, press , , and .
NOTE
If you measure a signal with a frequency outside the frequency range defined in
the frequency dependent offset table, the power meter uses the highest or lowest
frequency point in the frequency dependent offset table to calculate the offset