5 Using 8480 Series Power Sensors
160 Keysight N1913/1914A User’s Guide
Figure 5-3 Reference Calibration Factor pop-up window
5 Press to complete the entry.
Now, zero and calibrate the power meter/sensor combination as follows:
6 Press and the channel softkey. The Zeroing pop-up is displayed.
7 Connect the power sensor to the POWER REF output.
8 Press and the channel softkey to start the calibration routine. The
Calibrating pop-up is then displayed.
Now, set the sensor calibration factor for the frequency of the signal you want to
measure.
9 Check the current reference calibration factor setting by pressing ,
. The value is displayed on the Cal Fac field.
You can reduce the steps required to carry out the zero and calibration
procedure as follows:
– Connect the power sensor to the POWER REF output.
– Press and . (For dual channel meters, press
, or as required).
Does this setting match the value for the sensor? (The power sensor reference
calibration factor can normally be found above the calibration factors table on
the power sensor body.)