Maintenance 9
Keysight N1913/1914A User’s Guide 221
Instrument Self Test
If is selected, the following tests are run: (These are the same tests
which are run using the *TST? command.)
– Test Point Voltages
– Calibrator
–Fan
– RTC Battery
– Channel CW Path
As each test takes place, the name of the test is listed on the screen. While a test
is running, the message Testing... appears beside the name of the test. As each
stage of the test is completed, the Testing... message is replaced by either
Passed or Failed.
Figure 9-1 Self test complete
When the test is complete, the result will be displayed. Press to return to
the Service menu.
If the self test failed, information about the failure is displayed on the screen.