EasyManua.ls Logo

Keysight N1913 - Figure 5-4 Calibration Factor Pop-Up Window; Figure 5-5 Calibration Factor Display

Keysight N1913
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using 8480 Series Power Sensors 5
Keysight N1913/1914A User’s Guide 161
10 To change the settings, use the and keys to highlight the Cal Fac
value field and press to display the Cal Factor pop-up. Use the
numeric keypad to enter the required value in the Cal Factor pop-up window.
Figure 5-4 Calibration factor pop-up window
11 Press to complete the entry.
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
Figure 5-5 Calibration factor display
%
Calibration Factor

Table of Contents

Related product manuals