2 General Power Meter Functions
62 Keysight N1913/1914A User’s Guide
Figure 2-4 shows that how you can apply a Channel Offset or a Frequency
Dependent Offset prior to any mathematical functions. These allow you to
compensate each channel individually. An overall offset can be applied if required
using the Display Offset.
Setting Channel Offsets
This gain or loss is applied to the measured power before any mathematical
functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
A positive value compensates for a loss and a negative value compensates for a
gain.
Procedure
To enter a channel offset:
1 Press to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
2 Press to display the Offsets Setup.
3 Use the and keys to highlight the Offset setting field.
4 Press to check the Offset setting field.
Figure 2-5 Typical channel offset display
5 Press to highlight the Offset value field and press to display the
Offset pop-up. Use the numeric keypad to enter the required value in the
Offset pop-up window.
Offsets
Offset value field
Offset setting field