EasyManua.ls Logo

Keysight N1913 - Sensor Calibration Tables

Keysight N1913
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using N8480 Series Power Sensors 6
Keysight N1913/1914A User’s Guide 187
Sensor Calibration Tables
This section describes how to use sensor calibration tables. Sensor calibration
tables store the measurement calibration factors, for a power sensor model or for
a specific power sensor, in the power meter. They are used to correct
measurement results.
Use sensor calibration tables when you want to make power measurements over a
range of frequencies using one or more power sensors.
The N1913/1914A EPM Series power meters are capable of storing 20 sensor
calibration tables, each containing up to 80 frequency points. The power meter is
supplied with a set of nine predefined sensor calibration tables plus a “100%”
default table. The data in these tables is based on statistical averages for a range
of Keysight Technologies power sensors. Your own sensor will most likely differ
from the typical to some degree. If you require best accuracy, create a custom
table for each sensor you use as shown in “Editing/Generating Sensor Calibration
Tables on page 191.
To use calibration factor tables,
1 Select the sensor calibration table to be applied to a channel.
2 Zero and calibrate the power meter. The reference calibration factor used
during the calibration is automatically set by the power meter from the sensor
calibration table.
3 Specify the frequency of the signal you want to measure. The calibration factor
is automatically set by the power meter from the sensor calibration table.
4 Make the measurement.

Table of Contents

Related product manuals