EasyManua.ls Logo

Keysight N1913 - Figure 5-6 Sensor Table Selected; Figure 5-7 Frequency Dependent Offset Indicator

Keysight N1913
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Using 8480 Series Power Sensors
164 Keysight N1913/1914A User’s Guide
Procedure
First select the table for the sensor you are using as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and press
to highlight .
Figure 5-6 Sensor table selected
3 Press to complete the selection of the calibration factor table.
Figure 5-7 shows which offset table is selected.
Figure 5-7 Frequency dependent offset indicator
4 To change the frequency, press and use the and keys to
highlight the Frequency field.
Tables
Sensor Cal Tables
Table
On
NOTE
When no data is contained in the highlighted table, the key is disabled
(grayed out).
Table
Done
Sensor Table 5 Selected

Table of Contents

Related product manuals