5 Using 8480 Series Power Sensors
164 Keysight N1913/1914A User’s Guide
Procedure
First select the table for the sensor you are using as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and press
to highlight .
Figure 5-6 Sensor table selected
3 Press to complete the selection of the calibration factor table.
Figure 5-7 shows which offset table is selected.
Figure 5-7 Frequency dependent offset indicator
4 To change the frequency, press and use the and keys to
highlight the Frequency field.
When no data is contained in the highlighted table, the key is disabled
(grayed out).