EasyManua.ls Logo

Keysight N1913 - Selecting a Sensor Calibration Table; Figure 6-8 Sensor Table Selected

Keysight N1913
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Using N8480 Series Power Sensors
188 Keysight N1913/1914A User’s Guide
Selecting a Sensor Calibration Table
You can select a calibration factor table from the key menu followed by
and .
The State column indicates if any calibration factor tables are currently selected.
The Sensor Tbls screen is shown in Figure 6-8.
Procedure
Select sensor calibration table as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and press
to highlight .
Figure 6-8 Sensor table selected
Tables
Sensor Cal Table
NOTE
You can also view which sensor table is being used by pressing the ,
and use the and keys to highlight the CF Table setting field and
press to display the table.
Offset
Tables
Sensor Cal Tables
Table
On

Table of Contents

Related product manuals