Using U8480 Series USB Thermocouple Sensors 8
Keysight N1913/1914A User’s Guide 209
Power Meter Configuration
The N1913/1914A EPM Series power meters automatically recognize an U8480
Series USB thermocouple sensor when it is connected. The sensor calibration
data is automatically read by the power meter. The power meter also configures
the auto-averaging settings shown in Figure 8-1 to suit the thermocouple sensor
characteristics.
Figure 8-1 U8480 Series auto-averaging settings
NOTE
These values are valid only for the power meter channel connected with U8480
Series USB thermocouple sensors.
1234
4 64 256 512
Number of averages
2 8 128 256
2 2 16 256
22432
1128
Minimum power
Maximum power
Resolution setting
20 dBm
10 dBm
–10 dBm
–20 dBm
–30 dBm
1128
0 dBm
Dynamic range
–35 dBm