Maintenance 9
Keysight N1913/1914A User’s Guide 231
310
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp (temperature
compensation).
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor manual for details on
returning it for repair.
310
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance the power meter detects
power sensors connected to both it’s front and rear channel inputs.
–351 Configuration memory lost; storage fault
Refer to “Memory Erase/Secure Erase” on page 119
to perform secure erase.
321
Out of memory
The power meter required more memory than was available to run an internal operation.
330
Self-test Failed;
The -330, “Self-test Failed” errors indicate that you have a problem with your power meter. Refer to
“Contacting Keysight Technologies” on page 235 for details of what to do with your faulty power meter.
330
Self-test Failed;Measurement Channel Fault
330
Self-test Failed;Measurement Channel A Fault
330
Self-test Failed;Measurement Channel B Fault
330
Self-test Failed;Calibrator Fault
Refer to “Calibrator” on page 222 if you require a description of the calibrator test.
330
Self-test Failed;ROM Check Failed
330
Self-test Failed;RAM Check Failed
350
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
361
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be guaranteed.
362
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot be guaranteed.
363
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer contained data from
a previous command (the previous data is not overwritten). The output buffer is cleared when power has
been off, or after *RST (reset) command has been executed.