2 General Power Meter Functions
78 Keysight N1913/1914A User’s Guide
NOTE
Pulse power averages out any aberrations in the pulse such as overshooting or
ringing. For this reason, it is called pulse power and not peak power or peak
pulse power.
In order to ensure accurate pulse power readings, the input signal must be
pulsed with a rectangular pulse. Other pulse shapes (such as triangle, chirp, or
Gaussian) will cause erroneous results.
The pulse power on/off ratio must be much greater than the duty cycle ratio.