General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 85
10 Press or to setup measurement settings such as
measurement averaging, measurement frequency, offsets, duty cycle and so
forth. Refer to “General Power Meter Functions” on page 55 for the setup
procedures.
11 Set the frequency range and step by sending the below commands to meter
using remote interface.
SENSe1:FREQuency:STARt <start_frequency><frequency_unit>
SENSe1:FREQuency:STOP <stop_frequency><frequency_unit>
SENSe1:FREQuency:STEP <frequency_step_size>
12 Send *OPC (OPeration Complete) command to meter; to set the operation
complete bit in the Standard Event Status register when all pending device
operations are completed.
13 Set the meter to continuous trigger cycles by sending INITiate:CONTinuous
ON command to meter.
14 Configure and set the required power sweep range and step of the power
source.
15 Set the power source trigger input and trigger output.
16 Set the power source to sweep.
NOTE
Refer to the EPM Series Power Meters Programming Guide for the commands
usage details.
Example of commands set:
FREQ:STAR 10MHz
FREQ:STOP 500MHz
FREQ:STEP 10
NOTE
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488)
programming interfaces. Refer to N1913/1914A EPM Series Power Meters
Installation Guide for details.