LumaSMART
®
User Manual Introduction • 24
Operating Specifications
Table 2: LumaSMART Controller Specifications
Channels
Number of Channels 4 to 16 (in increments of 2)
Probes
Accepts all LumaSense DipTip Rugged Probe and Quality Probe
sensors
Environmental
Temperature Measurement Range -30 to +200 °C
Temperature Accuracy ±2 °C
Operating (Ambient)
Temperature
-30 to +70 °C
Operating Humidity 20 to 80% relative humidity, non-condensing
Storage Temperature -35 to +75 °C
Pollution Degree Degree 2: normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by
condensation must be expected.
Front Panel
Displays Temperature (°C)
Probe Signal Strength (%)
Status
Errors
Setup
Diagnostics
Update Interval Fixed at 10 seconds
System Power LED Green when power on
1 Form-C relay System Status relay
Power
Power Supply Universal AC: Phoenix Connector MC1.5/3-STF-3.81
(Part # 1827716) with Phoenix Ferrule AI0.75-8 GY (Part # 3200519)
Input Power 90 to 264 VAC or 127 to 370 VDC, 47 to 63 Hz
Wattage 108 W (maximum)