Research grade helium is recommended. If unavailable, use helium with a dew point of -67 °C
(-88 °F) or lower. Carbon dioxide, argon, dry air, or nitrogen can also be used for different applic-
ations (a multigas option is available for connection of multiple gases).
Gases
1 cm
3
chamber 1.14 cm ID × 1.1 cm D (0.45 in. ID × 0.44 in. D)
10 cm
3
chamber 1.85 cm ID × 3.95 cm D (0.72 in. ID × 1.55 in. D)
100 cm
3
chamber 4.62 cm ID × 6.18 cm D (1.82 in. ID × 2.43 in. D)
350 cm
3
chamber 5.84 cm ID× 13.94 cm D (2.30 in. ID × 5.49 in. D)
2000 cm
3
chamber 9.52 cm ID × 26.00 cm D (3.80 in. ID × 10.20 in. D)
Sample Cups
Precision
Reproducibility is typically to within ±0.01% of the nominal full-scale
sample chamber volume.
Reproducibility is guaranteed to within ±0.02% of the nominal full-
scale volume on clean, dry, thermally equilibrated samples using
helium in the 15 to 35 °C range.
Accuracy Accurate to within 0.03% of reading, plus 0.03% of sample capacity
Gas Consumption
Approximately 0.1 cm³ STP times the nominal cell volume for each
cycle
Approximately 2 cm³ STP times the nominal cell volume per ana-
lysis
Analysis
Specifications for AccuPyc II
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021
1 - 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals