E:Sense Error message: if no load cells are connected, see also
Chapter 16.1.
E:NoCom Error message: if there is no communication with the xB-
PI scale (see also Chapter 16.3).
E:HardE Error message: unable to read weight values from the
ADC (analog-digital converter); see also Chapter 16.1.
E:NotRd Error message: no load cells or no scale is connected (see
also Chapter 16.3).
The weight display appears.
Set the date and time after turning on the device for the rst time; see Chapter 7.15.2.
7.3 Switching o the device
The device is switched o/disconnected from power supply by pulling the plug.
7.4 Warm-up time
The device requires a warm-up time of 30 minutes before calibration.
7.5 Finding and connecting a device with a notebook/PC
If the device is connected to a notebook/PC via a point-to-point connection, an IP
address is negotiated via function "AutoIP". This can take up to 2 minutes!
NOTICE
When the IT/DHCP network cable is temporarily connected between the notebook/
PC and a device, the DHCP server is lost and the notebook/PC returns to the auto-IP
address within approx. two minutes!
Reason: The DHCP server/client relationship is checked cyclically at 23-minute
intervals.
1. On the notebook/PC, set the LAN local and Internet Protocol properties to "Obtain an
IP address automatically" depending on the operating system.
2. On the device, under - [Network parameter], activate the "Use DHCP" parameter
(default/factory setting).
7 Commissioning Transmitter in eld housing PR 5230
EN-109 Minebea Intec