EasyManua.ls Logo

Minebea Intec PR 5230 - Xbpi Calibration with Automatic Weight Detection

Minebea Intec PR 5230
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.13.8 xBPI calibration with automatic weight detection
Requirements:
- The xBPI protocol has been selected (see Chapter 7.13.2).
- The "xBPI-scale" weighing point has been selected (see Chapter 7.13.3).
- The platform has been set up (see Chapter 7.13.4).
- In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Conguration] -
[Weighing parameters] - [Conrming adjust.] was set to "manual".
- The communication between the device and platform is active.
Procedure:
The weight is displayed in high-resolution (10x).
7. Remove the weight.
8. Press the softkey to exit the menu and to save the settings.
1. Select -[Weighing point]- [xBPI-Scale] and conrm.
2. Press the [Setup] softkey.
The parameters of the xBPI-scale are read into the device.
3. Select [Calibration]- [Adjust with auto weight] with the cursor and conrm.
The weight is specied automatically.
The calibration process is carried out without a weight. The calibration status is
displayed.
7 Commissioning Transmitter in eld housing PR 5230
EN-159 Minebea Intec

Table of Contents

Related product manuals