[Zerotrack range]
Range within which automatic zero tracking compensates deviations.
Setting range: 0.2510000.00 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any
more.
Setting range for automatic tracking increments: 0.2510 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.125 s
At 0.0 s the zero tracking is switched o.
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum load (Max) without error message.
Setting range: 09999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Minimum weight]
Minimum weight at which a print command can be triggered.
Setting range: 09999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: <Single range>, Multiple range, Multi-interval
For scale range selection, see Chapter 7.12.15.2 and 7.12.15.3.
Press to exit the menu and to save the settings.
7.14.13 Subsequent dead load correction
If the hopper/platform weight changes by an amount that is higher than the zero-setting
range; e.g.due to dead load reduction, dead load increase, or mechanical changes, the
functions for automatic zero tracking and manual zero setting no longer work.
7 Commissioning Transmitter in eld housing PR 5230
EN-175 Minebea Intec