4.4.7 Digital inputs
The main board is equipped with 3 digital inputs for the process control. They are
electrically isolated by optocouplers, each bipolar potential-free.
The interface can be congured by software.
Depending on the order, the hardware is factory-set to "passive" (order code "DE1") or
"active" (order code "DE2").
Specications
Description Data
Connection Terminal, 6-pin
Number of inputs 3 (CH1, CH2, CH3)
Input supply voltage
Logic 0: 05 V DC or open
Logic 1: 1028 V DC
Active: Internal 12 V DC power supply.
Passive: External power supply required.
Input current ≤11 mA @ 24 V DC
≤5 mA @ 12 V DC
Protection against incorrect polarity.
Potential isolation Active:
Jointly supplied via potential-free voltage.
Passive:
Via optocoupler.
Cables Screened
Connect cable screen (wire gauge max. 1.5 mm
2
) to the sleeve
of the cable gland.
Cable length max. 50 m
4 Device installation Transmitter in eld housing PR 5230
EN-57 Minebea Intec