16.6 General error messages
Display VNC text Error and possible cause Remote dis-
play
Duplicate ID The IP address is already in use.
Two instruments with identical IP addresses.
Supply voltage is low! Supply voltage is too low -15%.
Supply voltage is too low (≤85%).
E:BadDev* Wrong serial number - Weighing electronics board has been chan-
ged after calibration.
- Device is not calibrated.
Error?
* Only if CAL switch is closed.
16.7 Error messages for Ex applications
In the event of load cell supply faults, the following error message is displayed:
E:Sense (via VNC: No sense voltage)
Two dierent faults are possible, which are displayed by two LEDs on the weighing
electronics board (WE1).
V120 is lit:
The sense voltage is not available or too low.
Cause:
Sense line or supply line is interrupted
16 Error messages Transmitter in eld housing PR 5230
EN-291 Minebea Intec