Note:
VNC text, see also Chapter 16.1.
LED E:Arith E:Invrs E:Overl E:NoSig E:Sense E:HardE/ E:NoCom
Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz Altern. ash. 1Hz
Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz
Flash. 1Hz Flash. 1Hz Flash. 1Hz Flash. 1Hz Altern. ash. 1Hz Altern. ash. 1Hz
Note:
With all other messages, the upper status indicator LED blinks.
16.3 Error messages for xBPI scales
Display VNC text Error and possible cause Remote dis-
play
E:NoSig Ext. meas. device error No weight values can be read from scale:
- Scale error
Error 3
E:Value Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
E:NoCom No values from scale No communication with xBPI scale:
- Cable break.
- Internal scale error.
- The scale is not connected to the power sup-
ply.
Error 9
E:NoWgt No weight data No weight display:
- Another weighing point was selected.
Press to assign the new weighing point to
the device.
Error •
16 Error messages Transmitter in eld housing PR 5230
EN-287 Minebea Intec