4.5.8.2 Connecting load cells with a 6-wire cable
The load cell cable screens are connected in the metal cable glands; see Chapter 4.2.4.
The 6 load cell cores are color-coded:
Terminal Port/color code Description
M+ + Meas./gn + Measuring voltage (load cell output)
M- - Meas./gy - Measuring voltage (load cell output)
+ Supply/rd + Supply voltageV+
+ Sense/wh + Sense voltage
- Supply/bu - Supply voltageV-
- Sense/bk - Sense voltage
Note:
The load cells are connected directly. Corner adjustment on the load cell junction board
is not provided.
4.5.9 Connecting an analog platform (series CAP…)
You can connect an analog platform to the device.
4 Device installation Transmitter in eld housing PR 5230
EN-71 Minebea Intec