7.15.9.3 Saving parameters
7.15.10 Conguring digital outputs
Congure the required function for [Output 1] to [Output 3] by selecting a signal from the
list.
The output is set to the corresponding state.
Selection list for output functions
Function SPM Bit Description
no condition --- No condition
actual diginp1 X00 = 0 digital input 1: not active
actual diginp2 X01 = 0 digital input 2: not active
actual diginp3 X02 = 0 digital input 3: not active
limit 1 out X16 = 0 Limiting signal 1: not active
limit 2 out X17 = 0 Limiting signal 2: not active
limit 3 out X18 = 0 Limiting signal 3: not active
ADC error X32 = 0 General error in weighing point: not ac-
tive (no error)
above Max X33 = 0 Weight above Max: not active
overload X34 = 0 Weight above Max plus the 'overload'
value: not active
below zero X35 = 0 Weight not below zero
center zero X36 = 0 Weight not within ¼ d of zero
inside ZSR X37 = 0 Weight not within zero-setting range
standstill X38 = 0 Standstill not active
out X39 = 0 Weight not below zero or above Max
command error X48 = 0 For internal use only.
command busy X49 = 0 For internal use only.
power fail X50 = 0 Set after power-on (=power failure): not
active
test active X56 = 0 Analog test was not started.
cal active X57 = 0 For internal use only.
tare active X58 = 0 Instrument is not tared.
marker bit 1 X64 = 0 Marker bit 1 not set, after power-on the
markers are set to "0".
2. Select and conrm the appropriate line (here: If input 1 changes from 0 to 1 [Input 1
on], a taring signal is triggered only, if the condition under [Condition] is met (limit 1
out = active).
3. If applicable, select additional conditions for the other parameter and conrm.
Press the softkey to exit the menu.
Transmitter in eld housing PR 5230 7 Commissioning
Minebea Intec EN-204