EasyManua.ls Logo

Minebea Intec PR 5230 - Terminating;Saving Calibration

Minebea Intec PR 5230
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.14.10.1 Checking the corner load (dead load)
Note:
For scale structures with containers pay attention to the following:
- For asymmetric scale structures a corner correction is not necessary.
- But for symmetric scale structures corner correction may be required.
After assignment and calibration, the load cell positions have been dened clearly.
7.14.10.2 Mechanical corner correction
A Mechanical corner correction has to be carried out, if the load is not evenly distributed
over the load cells, e.g. if the platform is wobbling.
The dead load on the load cells can be corrected using shims. If two coupled platforms are
connected, corner load checking or installation of shims for the platforms must be
performed independently.
A ne calibration can be done by software corner correction, see Chapter 7.14.10.3.
7.14.10.3 Software corner correction
If the corners are loaded in succession, the same value should be displayed on the device
at all times. An excessive deviation almost always means that the scale is tilted or
indicates load cell force shunts.
If the signal deviations cannot be resolved by carefully leveling the scale, the software
must be calibrated.
The menu is accessible via -[Weighing point]- [Weighing point A]- [Assign].
7.14.11 Terminating/saving calibration
The calibration is terminated by pressing the key.
1. Press the [Modify] softkey.
2. Select and conrm [Corner correction].
3. Set the CAL weight on an area of the scale structure.
The position(e.g.: LC 4) is highlighted.
4. Conrm this position.
Is displayed by ☑.
5. Remove the CAL weight.
6. Repeat steps 3 to 5 for the remaining load cells. You are free to choose any desired
order.
7. If all load cells have been loaded one time, press the [Calc] softkey to perform the
corner correction.
When corner correction is completed, it is marked with "OK".
The total weight remains unchanged. Only the eect of the individual load cells is
corrected.
8. Press to exit the menu and save.
Transmitter in eld housing PR 5230 7 Commissioning