Display VNC text Error and possible cause Remote dis-
play
E:NotRd Scale not ready The scale is not ready for weighing:
- The instrument is in the warm-up phase.
- The instrument is in automatic taring mode.
- The instrument was switched on with the
scale loaded.
Switch the device o and on again.
Error <
E:BadDev Wrong serial number Serial number of scale does not match the num-
ber set in the device.
Error?
16.4 Error messages for Pendeo® load cells
Display VNC text Error and possible cause Remote dis-
play
E:Overl Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
E:Value Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
E: Units Incompatible units Incompatible units:
- Incorrect calibration values e.g.due to incor-
rect calibration
Error 8
E:NoCom No values from scale No communication with Pendeo load cell:
- Cable break.
- Internal scale error.
- The junction box is not connected to the sup-
ply voltage.
Error 9
E:NotRd Scale not ready The scale is not ready for weighing:
- At least 1 load cell gives an error status or is
defective (no communication).
Error <
E: Cong Wrong conguration The number of load cells does not match the con-
guration.
Error =
E:BadDev Wrong serial number Serial number of scale does not match the num-
ber set in the device.
Error?
Transmitter in eld housing PR 5230 16 Error messages