Saving conguration and calibration data
Loading conguration and calibration data into the device
NOTICE
All data which can be congured in the Setup menu are overwritten!
If the le is loaded into several devices, changing the network settings and the host
name is indispensable!
9.1.4 Resetting the device to the factory settings
Note:
Reset to the factory settings is possible only, when the CAL switch is open. The IP
address and the Hostname remain unaected.
1. Click on [Backup] to create a backup .g. on the notebook.
2. Click on [Save as].
3. Create and open the required directory e.g. on the notebook.
4. Click on [Save] to save the backup le in the relevant directory.
1. Open the CAL switch; see Chapter 7.1.3.1.
2. Click on [Select File] (depending on the Internet browser).
3. e.g.on the notebook, navigate to the folder where the backup le was saved.
4. Select Backup le.
5. Click on [Restore].
The selected le is loaded into the device.
6. Click on the symbol in the Internet browser to return to the WEB menu.
7. Close the CAL switch.
1. Click on .
2. Click on .
The following prompt window appears:
9 Extended functions Transmitter in eld housing PR 5230