4.5.4 Connecting 28 load cells (650 ) using a 6-wire connecting cable
Connections are made via junction box PR 6130/.. using connection cable PR 6135/.. or
PR 6136/... .
Load cell supply circuit
- Load resistance of load cell circuit ≥75 , e.g. 8 load cells of 650 each
- Supply voltage is xed to 12 V DC and protected against short circuit.
For further technical data, see Chapter 4.4.3.
4.5.5 Connecting load cells of type series PR 6221
See installation manuals of PR 6221 and PR 6021/08, -/68.