7.12.3.2 Modifying a calibration
Note:
[Modify] may be used only for minor changes (e.g.changing the dead load, adapting
mV/V values for dead load and/or Max). Otherwise, always use [New]!
Example:
Resetting the Dead Load
3. By pressing [Continue] the data are reset to default rst (default) before performing
calibration. Press [Cancel] to cancel selection.
4. Determining the maximum load [Max], see Chapter 7.12.4.
5. Determining the scale interval [Scale interval], see Chapter 7.12.5.
6. Determining the dead load [Deadload at], see Chapter 7.12.6.
7. Calibrating with load [Max at], see Chapter 7.12.7.
8. Calibrating with mV/V [Max at], see Chapter 7.12.8.
9. Calibrating with load cell data (smart calibration) [Max at], see Chapter 7.12.8.1.
10. Carrying out linearization, see Chapter 7.12.11.
1. Select -[Weighing point]- [Calib] and conrm.
2. Press the [Modify] softkey.
7 Commissioning Transmitter in eld housing PR 5230
EN-127 Minebea Intec